Common Reasons for Sterilization Failure |
![]() |
|
Steam Autoclaves Interruption of Sterilization - Cycle Interruptions most frequently occur to retrieve a specifically required instrument. Effective sterilization in steam sterilizers is a specific function of time, temperature and pressure (saturated steam). If any of these parameters is compromised, sterilization failure is likely. Time - For most steam sterilizers running at 250 degrees Fahrenheit (121 degrees Celsius), a 20-minute cycle is adequate. That 20-minute cycle should begin at the time the sterilizer reaches operating temperature and pressure. So, it is important to allow adequate warm-up time (normally, 5-10 minutes) before starting the timing of the cycle. There are some steam sterilizers, commonly referred to as "rapid- cycle" or "flash" sterilizers, which may have a shorter cycle time (3.5 to 8 minute cycle). Temperature - Most steam sterilizers operate in a range between 250-270 degrees Fahrenheit. Any lower temperature may be inadequate to achieve sterilization. Pressure should be 15-30 psi (pounds per square inch). A frequent reason for loss of pressure is an inadequate seal of the sterilizer door gasket. The gasket may need to be cleaned with common dish washing detergent or it may need to be replaced. A "hissing" sound coming from the sterilizer door may be an indication that the gasket is not sealing properly. Venting is the process, which during the warm-up time, prior to the actual sterilization cycle, cool air is displaced within the sterilizer chamber by pressurized saturated steam. This process is achieved via a valve commonly referred to as a "bleeder valve". If this valve is not cleaned regularly, sterilization failure can occur. While the sterilizer temperature gauge may indicate that adequate sterilization temperature has been reached, there could be "cool" spots within the sterilizer chamber. The procedure for cleaning this valve should be described in the sterilizer owner's manual. Water - Only distilled or deionized water should be used in autoclaves for purposes of sterilization and maintenance related cleaning. Water, other than distilled water, may be too corrosive or cause a build up of mineral deposits that can impair the function of the sterilizer. Chemical Vapor Sterilizers (Chemiclaves®) Time - 20 minutes is the normal cycle time for chemical vapor sterilizers. Again, it is important to allow adequate warm-up time before starting the sterilization cycle. Temperature - Most chemical vapor sterilizers operate at the upper end of between 250-270 degrees Fahrenheit (121-131 degrees Celsius). Metering Valve - This valve on chemical vapor sterilizers is the device that meters the appropriate amount of chemical vapor solution that is introduced into the sterilizer chamber. If the valve is not maintained and cleaned regularly, sterilization failure can occur. The owner's manual should describe how to properly service this valve. Dry Heat Sterilizers Time - For most dry heat sterilizers a cycle time of 1 hour from the time the sterilizer reaches operating temperature is recommended. There are some dry heat sterilizers that have a shorter cycle time. Temperature - Most dry heat sterilizers operate at 350+ degrees Fahrenheit. Dentronix® - This brand of dry heat sterilizer operates via heat transfer. The racks within the sterilizer chamber get hot and, in turn, transfer heat to the instruments. For purposes of spore testing, it is necessary for the spore strip to be in direct contact with metal. It is recommended that the spore strip either be tied to a plier handle with ligature wire or paper clipped to a rack within the chamber. Some models of the Dentronix have a built-in holder for the bacterial spore strip. Newer Sterilizers HANSAmed Limited - your source for information on sterilizer monitoring. Contact us for questions regarding your sterilizer protocol, research or other related questions. |